Skip to main content
. 2018 Jul;82(3):208–215.

Table II.

Results of intraocular pressure measurements [median, min to max, mean ± standard deviation (SD)] and corresponding P-values in cattle, sheep, and goats using the TonoVet and Tono-Pen AVIA.

Eye TonoVet
Median (min to max)
Tono-Pen AVIA
Median (min to max)
TonoVet
Mean ± SD
Tono-Pen AVIA
Mean ± SD
Cattle OD 23 (15 to 37) 15.5 (8 to 27) 23.9 ± 5.0 15.5 ± 3.9
22 (12 to 40) 16 (9 to 25) 21.5 ± 6.3 15.6 ± 4.3
P-value OS* < 0.000
OD* < 0.000
Sheep OS 11 (8 to 20) 10 (5 to 18) 12.7 ± 3.0 9.8 ± 2.7
OD 10.5 (7 to 20) 10 (6 to 18) 11.7 ± 3.3 10.5 ± 2.4
P-value OS* < 0.000
OD* 0.009
Goats OS 22 (9 to 34) 13 (4 to 25) 21.6 ± 5.4 13.0 ± 4.3
OD 24 (11 to 37) 13 (6 to 25) 24.3 ± 5.6 14.1 ± 4.6
P-value OS* < 0.000
OD* < 0.000
*

Significant difference (Wilcoxon test, P ≤ 0.05).

OS — oculus sinister; OD — oculus dexter.