Table 3.

Performance of the GTFs and ITFs in terms of the errors associated with the estimation of central BP from diametric PVR waveforms in training data.

Training (N=50) RMSE [mmHg] SP Error [mmHg] PP Error [mmHg] SP & PP Error Norm [mmHg]
(a) GTF
Arm TLG GTF 4.39* 4.92* 4.35 6.57*
Arm TLS GTF 3.70 3.52 3.84 5.21
Ankle TL GTF 3.96 4.35 4.65 6.37
(b) ITF
f-ITF 3.83 5.37 5.02 7.36
p-ITF1 2.11 2.01 3.08 3.68
p-ITF2 2.43 2.18 3.31 3.96

Errors are root-mean-squared across all subjects. SP: systolic pressure. PP: pulse pressure. SP & PP Error Norm: Euclidean norm of SP and PP errors. *Significantly different from the arm TLS GTF (p < 0.05 with Holms-Bonferroni correction). Significantly different from the f-ITF (p < 0.05 with Holms-Bonferroni correction).