. 2018 Jul 13;15:108. doi: 10.1186/s12985-018-1015-6

Table 2.

Performance of the Euroimmun ZIKV IgM-ELISA kit

Performance according to the type of serum samples ZIKV IgM-ELISA result % (95% CI)
Sensitivity Positive Negative
 Acute-phase samples of RT-PCR-confirmed Zika casesa 0 14 0.0 (0.0–23.2)
 Convalescent-phase samples of RT-PCR-confirmed Zika cases 1 7 12.5 (0.3–52.7)
Specificity Negative Positive
 Acute-phase sample of RT-PCR-confirmed dengue cases 58 0 100.0 (93.8–100.0)
  Samples of DENV-1 infected cases 19 0 100.0 (82.4–100.0)
  Samples of DENV-2 infected cases 19 0 100.0 (82.4–100.0)
  Samples of DENV-4 infected cases 20 0 100.0 (83.2–100.0)
  Samples of primary dengue cases 13 0 100.0 (75.3–100.0)
  Samples of secondary dengue cases 45 0 100.0 (92.1–100.0)
 Convalescent-phase sample of RT-PCR-confirmed dengue cases 58 1 98.3 (90.9–100.0)
  Samples of DENV-1 infected cases 20 0 100.0 (83.2–100.0)
  Samples of DENV-2 infected cases 18 1 94.7 (74.0–99.9)
  Samples of DENV-4 infected cases 20 0 100.0 (83.2–100.0)
  Samples of primary dengue cases 12 1 92.3 (64.0–99.8)
  Samples of secondary dengue cases 46 0 100.0 (92.3–100.0)
 Blood donors samples 23 0 100.0 (85.2–100.0)

aAcute-phase serum samples were collected at a mean of 3 days post onset of symptoms. Thus, the ZIKV IgM-ELISA was not expected to be positive at that time