Skip to main content
. 2017 Nov 16;90(1080):20170350. doi: 10.1259/bjr.20170350

Table 3.

Comparison of the relative SI of each structure and relative contrast between conventional T1 weighted and synthetic MR images

Conventional T1 Synthetic T1 p-value
Relative SI
 Cartilage 17.2 ± 7.5 15.2 ± 6.1 0.370
 Bone marrow 12.1 ± 3.8 13.8 ± 3.1 0.075
 Muscle 13.3 ± 4.9 12.0 ± 3.8 0.418
Relative contrast
 Bone marrow to cartilage 7.3 ± 2.9 6.7 ± 2.0 0.624
 Bone marrow to muscle 7.5 ± 2.0 6.1 ± 1.6 0.011

SI, signal intensity.