Table 1.
Icare TA01i | Icare PRO | Icare HOME | Icare ic100 | |
---|---|---|---|---|
Released (year) | 2003 | 2010 | 2014 | 2016 |
Probe | Long | Short | Long | Long |
IOP display | ○ | ○ | Connecting PC | ○ |
Position sensor | – | ○ | ○ | ○ |
Sitting position | ○ | ○ | ○ | ○ |
Supine position | – | ⊚ | – | – |
Lateral position | ○ | ○ | ○ | ○ |
Self/home tonometry | – | – | ⊚ | – |
For children/handicapped patients | ||||
Sitting (awake) | ⊚ | ○ | – | ○ |
Supine (sleeping) | – | ⊚ | – | – |
Bland–Altman plots | ||||
Mean IOP difference between GAT | Slightly higher in younger patients | Close to GAT | Slightly lower measured by both self and others | Maybe slightly lower |
Reported width of 95% LOA to GAT (mmHg) | 8.93–11.38 | 5.2–9.4 | 7.31–14.59 | 10.84 |
Note: ○= appropriate to use. ⊚ = more appropriate to use.
Abbreviations: GAT, Goldmann applanation tonometer; IOP, intraocular pressure; LOA, limits of agreement; PC, personal computer.