Table 2.
Parameter | Analytical condition |
---|---|
Column | DB-WAX column (30 m × 0.25 mm, I.D. 0.25 μm film thickness) |
Column temperature | 0.75 min at 40 °C, increase to 60 °C at 10 °C/min, and increase to 150 °C at 3 °C/min and hold for 5 min |
Mass spectrometer | Ion source 200 °C Electron impact ionization potential 70 eV Select ion mode(m/z; 62, 74, 89) |
Mobile phase | He gas |
Flow rate | 0.9 mL/min |
Injection volume | 1 μL |
Injection mode | Splitless |