Table 2.

Accuracy and precision (intra and inter) of the method for the analysis of terephthalate metabolites, calculated from the analysis (n=5) of urine spiked at 4 different concentrations (0.5, 20, 40 and 100 ng/mL).

TPA mMTP mETP mTBTP mBzTP mEHTP

Spiked
(ng/mL)
Mean
(ng/mL)
RSD (%) Accur.
(%)
Mean
(ng/mL)
RSD
(%)
Accur.
(%)
Mean
(ng/mL)
RSD (%) Accur.
(%)
Mean
(ng/mL)
RSD (%) Accur.
(%)
Mean
(ng/mL)
RSD (%) Accur.
(%)
Mean
(ng/mL)
RSD
(%)
Accur.
(%)
Intra-day (n=5)
0.5 -* - - 0.47 2.7 94 0.58 2.1 116 0.56 4.3 112 0.43 5.1 86 -* - -
20 21.2 6.2 106 22.7 9.6 114 18.1 1.4 90.5 17.3 3.2 86.5 20.2 3.4 101 23.4 4.8 117
40 43.1 2.8 108 36.2 6.2 90.5 39.1 5.8 97.8 36.8 4.1 92 35.5 3.1 88.8 41.9 2.2 105
100 96.4 12.2 96.4 98.1 5.1 98.1 108 4.2 108 92.6 9.4 92.6 98.2 2.3 98.2 109 3.6 109
Inter-day (n=5)
0.5 -* - - 0.44 3.6 88 0.56 0.7 112 0.53 1.2 106 0.49 0.6 98 -* - -
20 20.8 3.7 104 21.9 4.7 110 18.8 5.4 94 18.2 3.9 91 19.1 2.7 95.5 22.1 3.7 111
40 41.5 2.1 104 36.4 5.9 91 40.2 6.6 101 38.5 5.4 96.3 36.9 6.4 92.3 43.7 6.2 109
100 94.6 8.4 94.6 97.5 11 97.5 104 1.9 104 88.7 6.9 88.7 101 3.5 101 105 12.1 105

LOD 0.85 0.12 0.05 0.03 0.04 1.25
LOQ 2.8 0.4 0.17 0.12 0.15 3.75

Where,

*

-not measured due to high background;

RSD: Relative standard deviation; Accur.: Accuracy (or recovery); LOD: Limit of detection; LOQ: Limit of quantification.