Table 1. Imaging protocol parameters of coronal T2WI, axial T1WI and T2WI, and axial high b-value DWI.

Parameters Coronal T2WI Axial T1WI Axial T2WI Axial high b-value DWI
Sequence TSE TSE TSE MS-EPI RESOLVE
Repetition time (ms) 3500 1490 3420 6200
Echo time (ms) 95 13 87 74/115
Fat suppression Dixon FS FS FS
Flip angle (°) 165 160 180 180
Section thickness (mm) 3 3 3 3
Section interval (mm) 0.3 0.3 0.3 0.3
Field of view (mm) 160 160 160 175
Matrix 224 × 320 256 × 320 256 × 320 174 × 174
Average 1 3 3 1, 1 and 4
Slices 24 24 24 24
Bandwidth (Hz/pixel) 300 250 260 685
Parallel acquisition factor 2 2 2 2
B-values (sec/mm2) / / / 0, 800, and 2000
Readout segments / / / 7
Acquisition time 2'13'' 2' 34'' 2' 39'' 14' 11''

Abbreviations: T2WI, T2-weighted imaging; T1WI, T1-weighted imaging; DWI, diffusion-weighted imaging; TSE, turbo spin-echo; MS-EPI, multi-shot echo-planar imaging; RESOLVE, readout segmentation of long variable echo-trains; FS, frequency-selective.