Table 4.
Paired Comparisons of subharmonic aided pressure estimation (SHAPE) in left ventricle (LV) and the manometer pressures for all canines.
| Difference (mmHg) (SHAPE║ – manometer pressure
|
95% Confidence Interval of the Difference (mmHg)
|
Significance (2-tailed) | ||||
|---|---|---|---|---|---|---|
| Mean | Standard deviation | Standard error of mean | Lower | Upper | ||
| M-LVDP* | 1.26 | 1.49 | 0.75 | −1.11 | 3.64 | 0.18 |
| LVDmin† | −0.03 | 1.81 | 0.91 | −2.92 | 2.85 | 0.97 |
| LVEDP‡ | 1.86 | 6.16 | 3.08 | −7.94 | 11.67 | 0.59 |
| LVPSP§ | 1.37 | 1.59 | 0.79 | −1.16 | 3.90 | 0.18 |
mean LV diastolic pressure,
minimum LV diastolic pressure,
LV end diastolic pressure,
LV peak systolic pressure