Formulations used to design cortisol-selective polymers.

The values are calculated by taking the ratio of slope of each sensor calibration curve for each molecularly imprinted and nonimprinted control polymer (Fig. 3, B and C, and figs. S1 to S5). The starting concentration of cortisol is 0.2 mmol in each condition. Control polymers are not included in the table; they were prepared exactly as corresponding selective polymer but without the template. AIBN, azobisisobutyronitrile.

Membrane Monomer Cross-linker Initiator Solvent Ratio* Conditions Sensing factor
A1618 MAA EDMA AIBN Acetonitrile 1:6:18 UV, 4°C, 30 hours 4.2
M1618 MAA EDMA AIBN Methanol 1:6:18 UV, 4°C, 30 hours 5.1
D1618 MAA EDMA AIBN DCM 1:6:18 UV, 4°C, 30 hours 5.4
D1630 MAA EDMA AIBN DCM 1:6:30 UV, 4°C, 30 hours 10.1
D1660 MAA EDMA AIBN DCM 1:6:60 UV, 4°C, 30 hours 2.2
D1630RT MAA EDMA AIBN DCM 1:6:30 UV, room
temperature,
30 hours
6.7

*Molar ratio of template/functional monomer/cross-linker.

†Calculated based on each sensor device’s performance to express imprinting efficiency.