Table 4.

Transducer Reproducibility Estimates for Single- and Three-image AC and BSC Measures Obtained under Various Conditions, Calculated Using the Two-way Random Effects Model.

Conditions Summary Statistics for AC (Unit: dB/cm-MHz)

Participant Group # Images Averaged Mean Range Between-participant SD (95% CI) QIBA-reproducibility SD (95% CI) Gauge-reproducibility SD Between-trial Repeatability SD (95% CI) RDC (95% CI)
A 1
3
0.95 [0.66, 1.23] 0.11 (0.07, 0.16) 0.08 (0.07, 0.62) 0.03 0.08 (0.06, 0.10) 0.23 (0.20, 1.70)
0.96 [0.67, 1.21] 0.10 (0.07, 0.16) 0.07 (0.06, 0.55) 0.05 0.06 (0.05, 0.07) 0.21 (0.17, 1.52)

B 1
3
0.93 [0.56, 1.25] 0.13 (0.09, 0.19) 0.07 (0.06, 0.08) 0 0.07 (0.06, 0.09) 0.18 (0.16, 0.23)
0.93 [0.56, 1.24] 0.13 (0.09, 0.18) 0.06 (0.05, 0.09) 0 0.06 (0.05, 0.07) 0.16 (0.14, 0.26)

Conditions Summary Statistics for BSC (Unit: dB)

Participant Group # Images Averaged Mean Range Between-participant SD (95% CI) QIBA-reproducibility SD (95% CI) Gauge-reproducibility SD Between-trial Repeatability SD (95% CI) RDC (95% CI)

A 1
3
−27.36 [−39.23, −14.63] 3.56 (2.50, 5.39) 2.91 (2.51, 36.6) 1.09 2.69 (2.21, 3.45) 8.05 (6.94, 101)
−27.50 [−37.75, −17.89] 3.46 (2.42, 5.26) 2.56 (2.18, 30.2) 1.51 2.07 (1.70, 2.64) 7.09 (6.04, 84.7)

B 1
3
−26.39 [−38.54, −15.91] 4.63 (3.45, 6.81) 2.90 (2.49, 41.9) 0.78 2.79 (2.30, 3.55) 8.02 (6.90, 116)
−26.48 [−39.88, −15.14] 4.58 (3.42, 6.70) 2.68 (2.30, 39.4) 0.63 2.60 (2.15, 3.31) 7.42 (6.38, 109)