Table 2. Diagnostic performance of AFIAS for detecting HBsAg, anti-HBs, and anti-HCV.

Sample N TP FN TN FP Sensitivity % (95% CI) Specificity % (95% CI)
HBsAg 1,500 399 1 1,092 8 99.8 (98.6–100.0) 99.3 (98.6–99.69)
Anti-HBs 500 200 0 300 0 100.0 (98.2–100.0) 100.0 (98.8–100.0)
Anti-HCV 1,500 395 5 1,090 10 98.8 (97.1–99.6) 99.1 (98.3–99.6)

Abbreviations: TP, true positive; FN, false negative; TN, true negative; FP, false positive; CI, confidence interval; HBsAg, hepatitis B surface antigen; anti-HBs, antibody to HBsAg; anti-HCV, antibody to hepatitis C virus.