. 2018 Jul 26;56(8):e00214-18. doi: 10.1128/JCM.00214-18

TABLE 1.

Summary of RPR screen concordance between AIX1000 system and ASI carda

AIX1000 parameter ASI card value(s)
RX (%) NR (%) Total (%) % PPA % NPA K (95% CI)
Retrospective specimens
    RX 48 (48) 2 (2) 50 (50)
    NR 2 (2) 48 (48) 50 (50)
    Total 50 (50) 50 (50) 100 (100) 96.0 96.0 0.920 (0.843–0.997)
Prospective specimens
    RX 20 (1.9) 11 (1.1) 31 (3.0)
    NR 1 (0.1) 996 (96.9) 997 (97.0)
    Total 21 (2.0) 1,007 (98.0) 1,028 95.2 98.8 0.763 (0.634–0.893)
a

CI, confidence interval; Κ, kappa coefficient; NPA, negative percent agreement; NR, nonreactive; PPA, positive percent agreement; RX, reactive.