Discordant prospective specimen summarya

Specimen ASI card
AIX1000
Treponemal antibody status Interpretation
Screen Titer Screen Titer
464 NR RX 1:4 RX ASI card false negative
471 NR RX 1:2 RX ASI card false negative
613 NR RX 1:1 NR AIX false positive
629 NR RX 1:8 RX ASI card false negative
701 NR RX 1:1 RX ASI card false negative
850 NR RX 1:1 RX AIX false positive
1002 NR RX 1:1 NR AIX false positive
1009 NR RX 1:1 RX ASI card false negative
1205 RX 1:4 RX 1:1 NR Dual false positive
1241 NR RX 1:1 NR AIX false positive
1247 NR RX 1:1 NR AIX false positive
1300 RX 1:1 NR NR ASI card false positive
1412 NR RX UNDb RX ASI card false negative
a

NR, nonreactive; RX, reactive. Treponemal antibody status was defined as detection of a TP-PA RX result plus 1 additional RX treponeme-specific chemiluminescent immunoassay result. (See the “Traditional testing algorithm” section in Materials and Methods.)

b

UND, unable to determine (insufficient specimen volume for titer).