Comparing study between the present work with some recent impedimetric label—free immonosensors for E. coli O157:H7 detection.

Schematic Immunosensor Assembly Sensitivity [Ω mL/cfu*cm2] Linear Range [cfu/mL] LOD [cfu/mL] RSD [%] References
Au + PANI + Glu + anti-E.coli 1 × 102–1 × 107 100 [14]
Au + MHDA + anti-E.coli 772,050.00 3 × 101–3 × 104 20 2.0 [15]
Au-MBA-Ab 784,615.38 3 × 101–3 × 105 30 4.3 This work
Au-MBA-ProteinA/G-Ab 2,228,076.92 3 × 102–3 × 107 300 5.1 This work
Au-Cys-Ab 283,589.74 3 × 101–3 × 107 30 2.9 This work
Au-Cys-Ferrocene-Ab 141,923.07 3 × 100–3 × 105 3 3.8 This work
Au-Cys-PAMAM-Ferrocene-Ab 1,166,666.67 3 × 103–3 × 106 3000 4.7 This work

PANI: Polyaniline; Glu: Glutaraldheyde]; MHDA: 16-mercaptohexadecanoic acid; Cys: Cysteamine; PAMAM: Polyamidoamine dendrimer; Ab: antibody.