Table 1.

The size of gastrojejunostomy in each group.

Expected size Actual size The rate of change, %
MCT-A Cuboid, D = 5.07 ± 0.11 mm,
n = 10 Sa = 20.88 mm2 Sb = 20.17 ± 0.91 mm2 −3.40c
MCT-B Ring, D = 6.05 ± 0.08 mm,
n = 9 Sd = 28.26 mm2 Se = 28.74 ± 0.78 mm2 +1.70f
Hand-sewn D = 6.50, D = 7.06 ± 1.35 mm,
n = 7 Sg = 33.17 mm2 Sh = 40.38 ± 14.42 mm2 +21.7i

aMCT-A: The length, width and height of the cuboid magnets was 7.20 mm, 2.90 mm, 1.60 mm, respectively. The anastomosis size of expected was 20.88 mm2;

bMCT-A: The diameter of anastomotic was 5.07 ± 0.11 mm, so the actual anastomosis size was 20.17 ± 0.91 mm2;

cMCT-A: The rate of change = Sb–Sa/Sa;

dMCT-B: The outside diameter of the magnetic ring was 6 mm, so the anastomosis size of expected was 28.26 mm2;

eMCT-B: The diameter of anastomotic was 6.05 ± 0.08 mm, so the actual anastomosis size was 28.74 ± 0.78 mm2;

fMCT-B: The rate of change = Se–Sd/Sd;

gHand-sewn: The diameter of rats’ jejunum was approximately 6.5 mm, and the cross-sectional area of the jejunum is taken as the expected size of the anastomosis. The anastomosis size of expected was 33.17 mm2;

hHand-sewn: The diameter of anastomotic was 7.06 ± 1.35 mm, so the actual anastomosis size was 40.38 ± 14.42 mm2;

iThe rate of change = Sh–Sg/Sg.