Skip to main content
. 2018 Jul 13;201(4):1315–1326. doi: 10.4049/jimmunol.1800323

Table I. Repeatability of ka, kd, and Kd measurements of CSP Ag (at high density) binding of anti-CSP mAbs in PBS buffer and 1:50-diluted NHS measured in triplicate.

Ag Ab Diluent Mean ka (×105 M−1 s−1) Mean kd (×10−4 s−1) Mean Kd (nM)
PF16 AB236 PBS 1.82 ± 0.08 (%CV = 4.4) 1.98 ± 0.29 (%CV = 14.6) 1.09 ± 0.19 (%CV = 17.4)
PF16 AB236 NHS 2.20 ± 0.02 (%CV = 0.9) 3.90 ± 0.19 (%CV = 4.9) 1.77 ± 0.08 (%CV = 4.5)
NANP6 AB334 PBS 0.817 ± 0.002 (%CV = 0.2) <0.01a <0.01a
NANP6 AB334 NHS 1.02 ± 0.01 (%CV = 1.0) <0.01a <0.01a
CSP-FL AB236 PBS 4.62 ± 0.10 (%CV = 2.2) <0.01a <0.01a
CSP-FL AB236 NHS 6.33 ± 0.91 (%CV = 14.4) <0.01a <0.01a
CSP-FL AB334 PBS 0.57 ± 0.03 (%CV = 5.3) <0.01a <0.01a
CSP-FL AB334 NHS 0.37 ± 0.04 (%CV = 10.8) <0.01a <0.01a
a

SD and %CV were ND because of slower than resolvable dissociation rate (kd < 0.01 × 10−4 s−1).