Table 6.

Flow measurements in the post-surgery in vitro model run at 1 and 2 L/min. Vessel numbers denote the vessel locations shown in Figure 9b.

Post-Surgery Model
Vessel Volume Flow Rate (L/min)
4D Flow CFD Flow Probe
1L/min
1 0.467 0.470 0.470
2 0.449 0.460 0.460
3 0.341 0.374 0.370
4 0.283 0.273 0.270
5 0.282 0.283 0.280
2 L/min
1 1.027 0.970 0.970
2 0.948 0.910 0.910
3 0.742 0.756 0.740
4 0.591 0.552 0.540
5 0.626 0.572 0.560