Skip to main content
. 2018 Jul 15;2018:8568356. doi: 10.1155/2018/8568356

Table 2.

Invasive and noninvasive hemodynamic data from CardioMEMS implant and simultaneous echocardiographic assessment.

N=17
Invasive hemodynamics at CardioMEMS implant
 PAsP, mmHg (mean ± SD) 49 ± 16
 PAdP, mmHg (mean ± SD) 21 ± 7

Hemodynamics by CardioMEMS
 PAsP, mmHg (mean ± SD) 42 ± 8
 PAdP, mmHg (mean ± SD) 18 ± 4

Estimated hemodynamics by echocardiogram
 PAsP, mmHg (mean ± SD) 42 ± 8
 LAP, mmHg (mean ± SD) 19 ± 5

LAP, left atrial pressure; mPAP, mean pulmonary artery pressure; PAdP, pulmonary artery diastolic pressure; PAsP, pulmonary artery systolic pressure.