Comparison of Simulated and Measured Electric Current for Each Pulse Sequence.

Sequence Electrode Pair Average Measurements [A] Average Simulation [A] RMSE [A] Difference of the First and Last Pulse [A] Stationary Simulation [A]
1 2-6 48.6 46.7 1.9 1.2 48.3
2 1-6 46.2 50.5 4.3 −0.2 52.3
3 3-5 48.8 41.8 7 0.9 44.3
4 4-5 48.3 40 8.4 1.2 42.0
5 5-6 48.8 59.2 10.4 1.7 62.1
6 3-6 49.1 53.2 4.1 0.7 53.7
7 2-5 49.1 53 3.9 1.8 52.5
8 4-6 49.1 38.1 11.1 1.3 39.8
9 2-5 45.5 46 3.7 0.8 45.6
10 4-6 41.3 32.8 8.5 0.7 34.1
11 1-5 32.6 44 11.5 1.6 42.3

Abbreviation: A, Ampere; RMSE, root mean square of error.