Table 2. Correlation coefficients (with significance levels) between ORA/CST parameters against IOP measured by the Goldmann applanation tonometer, age and gender.
Parameters | GAT-IOP (mm Hg) | Patients' age | Patient's gender |
OD CH | 0.257 (0.014)a | -0.103 (0.334) | 0.093 (0.382) |
OD CRF | 0.381 (<0.001) a | -0.060 (0.573) | 0.087 (0.416) |
OD IOP-G | 0.379 (<0.001) a | 0.040 (0.705) | 0.004 (0.971) |
OD IOP-CC | 0.193 (0.068) | 0.107 (0.314) | 0.006 (0.955) |
OD IOP-CST | 0.243 (0.021) a | 0.145 (0.172) | -0.024(0.821) |
OD CCT-CST | -0.040 (0.705) | -0.066 (0.536) | -0.219 (0.038) a |
OD HCDA | -0.150 (0.159) | 0.060 (0.577) | 0.019 (0.859) |
OD A1L | -0.064 (0.549) | 0.024 (0.824) | -0.216 (0.041) a |
OD A2L | 0.084 (0.430) | -0.212(0.045) a | 0.181 (0.087) |
OD A1V | -0.189 (0.074) | -0.116 (0.275) | 0.171 (0.108) |
OD A2V | -0.027 (0.808) | -0.008 (0.940) | 0.162 (0.141) |
OD PD | -0.032 (0.764) | 0.145 (0.173) | 0.033 (0.759) |
OD RoC | 0.041 (0.701) | 0.043 (0.686) | -0.187 (0.077) |
OS CH | 0.062 (0.560) | -0.006 (0.958) | -0.051 (0.636) |
OS CRF | 0.216 (0.041) a | 0.043 (0.686) | -0.032 (0.762) |
OS IOP-G | 0.335 (0.001) a | 0.098 (0.357) | -0.078 (0.464) |
OS IOP-CC | 0.265 (0.012) a | 0.095 (0.372) | -0.060 (0.577) |
OS IOP-CST | 0.361 (<0.001) a | 0.001 (0.991) | 0.088 (0.407) |
OS CCT-CST | 0.111 (0.301) | 0.039 (0.714) | -0.240 (0.023) a |
OS HCDA | -0.426 (<0.001) a | 0.065 (0.543) | -0.107 (0.316) |
OS A1L | 0.119 (0.262) | -0.041 (0.703) | -0.141 (0.185) |
OS A2L | 0.214 (0.043) a | -0.069 (0.519) | -0.105 (0.325) |
OS A1V | -0.072 (0.499) | 0.062 (0.562) | 0.122 (0.253) |
OS A2V | 0.057 (0.602) | -0.002 (0.989) | 0.080 (0.463) |
OS PD | -0.235 (0.026) a | -0.053 (0.618) | -0.023 (0.827) |
OS RoC | 0.266 (0.011) a | -0.077 (0.471) | -0.013 (0.900) |
CST: Corvis ST; ORA: Ocular Response Analyzer; GAT-IOP: IOP measured by Goldmann applanation tonometry; OD: Right eye; OS: Left eye; CH: Corneal hysteresis; CRF: Corneal resistance factor; IOP-G: Goldmann-correlated IOP; IOP-CC: Corneal compensated IOP; IOP-CST: Corrected IOP measured by CST; CCT-CST: Central corneal thickness measured by CST; HCDA: Highest concavity deformation amplitude; A1L: A1 length; A2L: A2 length; A1V: A1 velocity; A2V: A2 velocity; RoC: Radius of curvature; PD: Peak distance. aP values less than 0.05.
r (P)