Table 2.
Sensitivity and LOD in Site 1 (GOx only) and Site 2 (GOx + GABAse).
| GABA concentration (μM) | Sensitivity (nA μM-1cm-2) | LOD (μM) | ||
|---|---|---|---|---|
| Site-1 | Site-2 | Site-1 | Site-2 | |
| 0 | 90 ± 5∗ | 132 ± 13 | 0.27 ± 0.02 | 0.08 ± 0.008 |
| 5 | 104 ± 8 | 146 ± 16 | 0.23 ± 0.01 | 0.08 ± 0.007∗ |
| 10 | 106 ± 10 | 154 ± 19 | 0.22 ± 0.01 | 0.07 ± 0.008 |
| 20 | 104 ± 12 | 164 ± 21 | 0.23 ± 0.02 | 0.07 ± 0.009 |
Error value is shown as mean ± SEM. Two-tailed Students t-test was performed (n = 6, p < 0.0001, ∗p < 0.05). One-way ANOVA was performed (n = 6, p < 0.0001) to verify that sensor-to sensor variation (in the same site) is not significant.