Skip to main content
. Author manuscript; available in PMC: 2019 Aug 1.
Published in final edited form as: J Neural Eng. 2018 Aug 1;15(4):046032. doi: 10.1088/1741-2552/aac810

Table 1.

The WOENI specifications

ECoG recording Optical stimulation
# of channels 2 # of channels 4
Mid-band gain 49 dB Stim. current 5~30 mA, 8 bits
Bandwidth 1~200 Hz Light intensity 4~12 mW/mm2
Input-referred noise 8.47 µVrms, 8.64 µVrms Stim. period 0.1~0.9 s in 10 steps, 1~10 s in 10 steps
THD (11.3 mVPP input) 1% Duty cycle 0.1~0.9% in 10 steps, 1~10% in 10 steps
Dynamic range (1% THD) 53.5 dB, 53.3 dB Data transmission
Input impedance @ 1 kHz 10 MΩ Data throughput 32 kbps
CMRR >110 dB @ 1 kHz Dara error rate < 0.3%
PSRR >80 dB @ 1 kHz Transmission range ~3 m
Power consumption 1.5 mW/channel ENOB 11 bits
Recovery time 5 ms Sampling rate 1 kS/s