Table 2.

Comparison of True Versus Measured Peak Systolic Velocity (PSV) by Carotid Artery Segment.

Common Carotid Internal Carotid External Carotid Pa
Number of measurements 16 10 10
Mean absolute deviation % 6 ± 6 9 ± 6 8 ± 4 NS
True PSV, cm/s 104 ± 34 115 ± 49 120 ± 54
Measured PSV, cm/s 98 ± 35 108 ± 41 125 ± 62
Pb .001 NS NS

Abbreviation: NS, not significant.

a

Analysis of variance, difference between artery segments.

b

Paired t test, true versus measured velocity.