Skip to main content
. 2018 Jun 28;23(7):1568. doi: 10.3390/molecules23071568

Table 1.

Calibration curves, limits of detection (LODs), and limits of quantification (LOQs) of the six analytes.

Analyte Linear Equation a R 2 b Linear Range
(μg/mL)
LOD (μg/mL) c LOQ (μg/mL) d
Bisdemethoxycurcumin y = 101,215x − 75,414 0.9999 3.13–100 0.20 0.67
Demethoxycurcumin y = 87,918x − 66,314 1.0000 1.56–100 0.26 0.88
Curcumin y = 86,839x − 64,613 0.9999 1.56–100 0.31 1.04
ar-Turmerone y = 26,801x − 33,652 1.0000 6.25–400 0.73 2.45
β-Turmerone y = 69,449x − 49,841 1.0000 3.13–200 0.51 1.71
α-Turmerone y = 35,041x − 46,310 1.0000 3.13–200 0.91 3.02

ay, peak area; x, concentration of the analytes (μg/mL); b the correlation coefficient; c limit of detection (S/N = 3); d limit of quantification (S/N = 10).