Table 4.

Summary of water HU accuracy and uniformity measured on corresponding the central slice, for different diameter cylinders, compared to ACR [12] and Hobson [16] recommendations

Metric ACR (CT) Recommendations Hobson et al. (CBCT) Recommendations Phantom Diameter (cm) AZOR (HU) saddle (HU)
HU Accuracy ±7 HU ±16 HU 15 9 −4
12.5 2 −1
10 1 0
HU Uniformity
(Center to Edge)
±5 HU ±22 HU 15 ±26 ±27
12.5 ±19 ±16
10 ±13 ±13