Validation parameters of all trans MK-7 in oil dietary supplements.

DAD Detector QTOF Detector (SIM, m/z = 666.5 ± 0.5 m/z)
LOD (ng mL−1) 290.3 200.9
% CV 9.1 9.8
LOQ (ng mL−1) 957.9 603.7
% CV 0.3 9.7
Recovery:
Spiked level 1: 10 µg mL−1 (%) 85.6 81.4
SD 1.7 6.1
Spiked level 2: 20 µg mL−1 (%) 89.3 90.5
SD 2.3 6.2
Spiked level 3: 30 µg mL−1 (%) 93.8 94.6
SD 0.9 4.7
Matrix effect (ME) (%) <5 <5
SD - -
Linear (range in µg mL−1 and R2) 0.9–250 µg mL−1 R2 = 0.997 Very low, for example: 2–4 µg mL−1 R2 = 0.991, but for 2–20 µg mL−1 R2 = 0.860
Precision and accuracy:
Intra-day precision in % CV 0.8 6.1
Intra-day accuracy 2.9 5.8
Inter-day precision in % CV 4.1 11.3
Inter-day accuracy −2.9 −5.0