Skip to main content
. 2018 Jan 31;8(4):271–276. doi: 10.1016/j.jpha.2018.01.003

Table 1.

Extraction recovery and matrix factor for tadalafil from human plasma.

Quality control level (ng/mL) Mean area response (n = 6)
Recovery (%)
Matrix factor
A (post-extraction spiking) B (pre-extraction spiking) C (neat samples in mobile phase) Analyte (B/A) IS Analyte (A/C) IS IS-normalized (Analyte/IS)
400 5,452,803 5,447,895 5,558,413 99.91 99.84 0.981 0.976 1.005
200 2,762,572 2,758,151 2,833,407 99.84 100.05 0.975 0.981 0.994
30.0 418,560 421,113 436,910 100.61 99.63 0.958 0.972 0.986
1.50 20,735 20,689 21,895 99.78 101.52 0.947 0.928 1.020
0.50 6816 6744 7172 98.95 99.58 0.953 0.950 1.003

IS: internal standard.