Skip to main content
. 2018 Aug 8;2018:7170539. doi: 10.1155/2018/7170539

Table 5.

Obtained results from validation of the proposed method: robustness.

Simvastatin Res E/F Res G/SIM Res B/C
Rt (min) T NTP
Flow (mL/min)
 0.6 mL/min 14.05 0.99 109188 1.52 4.48 1.66
 0.8 mL/min 13.78 1.08 128239 1.50 4.46 1.53
Content of acetonitrile in initial phase
 35% 14.77 1.05 234052 1.53 4.05 1.61
 45% 14.74 1.02 236896 1.58 4.17 1.66
Column temperature
 30°C 13.42 1.01 93614 1.51 4.68 1.89
 40°C 13.83 1.02 109994 1.53 4.43 1.78
Another column 15.07 1.05 169635 1.59 4.47 2.15