Table 1.

Relevant imaging parameters for the dMRI protocols investigated in the current study

TR (ms) MB factor b values (s/mm2) # of diffusion directions/shell # of b0 images # of slices Slice orient inter-slice shift (FOV) time shift (us)
Pilot study 7400 2 1500 65 8 144 Sagittal 1/2 1920
HCP 1Tx 7000 2 1000/2000 64 15 132 Oblique axial 1/2 1920
HCP pTx 7400 2 1000/2000 64 15 144 Sagittal 1/2 1920
pTx MB3 4934 3 1000/2000 95 22 144 Sagittal 1/4 960
pTx MB4 3701 4 1000/2000 129 29 144 Sagittal 1/4 640

Other relevant imaging parameters that were kept constant for all protocols listed include 1.05 mm isotropic resolution, 71 ms echo time, 3-fold in-plane acceleration (iPAT3), 6/8 partial Fourier, 210(RO)×210(PE) mm2 field of view, 0.82 ms echo spacing, and 1388 Hz/pixel bandwidth. For all of the HCP and pTx related acquisitions, the listed diffusion weighted and b0 images were acquired twice with inverted phase encode directions (ie Anterior-Posterior (AP) and PA) and the total acquisition was divided into 4 segments of ~10 minutes each. For the pilot study, the listed diffusion weighted and b0 images were acquired once with the AP phase encode direction, and another set of six b0 images were acquired with matched imaging parameters but with the PA phase encode direction.