Table 1.
Duodenoscope type | N | AM20 | MGO | ||
Contam. | Not contam. | Contam. | Not contam. | ||
All duodenoscopes | 150 | 33 (22%) | 117 (78%) | 23 (15%) | 127 (85%) |
Olympus TJF-Q180V | 69 | 15 (22%) | 54 (78%) | 15 (22%) | 54 (78%) |
Olympus TJF-160VR | 43 | 13 (30%) | 30 (70%) | 6 (14%) | 37 (86%) |
Olympus TJF-160R | 8 | 1 (13%) | 7 (87%) | 0 | 8 |
Olympus TJF-140R | 2 | 0 | 2 | 0 | 2 |
Olympus TJF-145 | 2 | 0 | 2 | 0 | 2 |
Pentax ED34-i10T | 11 | 3 (27%) | 8 (73%) | 0 | 11 |
Pentax ED-3490TK | 8 | 0 | 8 | 0 | 8 |
Pentax ED-3680TK | 1 | 0 | 1 | 1 (100%) | 0 |
Fujifilm ED-530XT8 | 5 | 0 | 5 | 0 | 5 |
Fujifilm ED-530XT | 1 | 1 (100%) | 0 | 1 (100%) | 0 |
Sample site | AM20 | MGO | |||
Contam. | Not contam. | Contam. | Not contam. | ||
All sample sites | 701* | 47 (7%) | 654 (93%) | 35 (5%) | 666 (95%) |
Biopsy channel | 146 | 5 (3%) | 141 (97%) | 6 (4%) | 140 (96%) |
Suction channel | 137 | 4 (3%) | 133 (97%) | 5 (4%) | 132 (96%) |
Forceps elevator | 148 | 14 (10%) | 134 (90%) | 7 (5%) | 141 (95%) |
Brush | 139 | 17 (12%) | 122 (89%) | 14 (10%) | 125 (90%) |
Protection cap | 56 | 6 (11%) | 50 (89%) | 3 (5%) | 53 (95%) |
Elevator channel | 53 | 0 | 53 | 0 | 53 |
Air/water channel | 26 | 1 (5%) | 21 (95%) | 0 | 22 |
*Sampling of all possible sites would have yielded 745 samples: 44 (6%) sites were not sampled. This included 4/150 (3%) biopsy channel, 13/150 (9%) suction channel, 2/150 (1%) forceps elevator, 11/150 (7%) brush, 9/65 (13%) protection cap, 2/55 (4%) elevator channel and 3/25 (12%) air/water channel samples.
AM20, microbial growth with ≥20 CFU/20 mL of any type of microorganism; Contam., contaminated; MGO, presence of any microbial growth of gastrointestinal or oral microorganisms; Not contam., not contaminated.