Skip to main content
. 2018 Jun 1;20(6):395–402. doi: 10.1089/dia.2018.0150

Table 5.

Continuous Glucose Monitoring Performance at Various Rates of Change (n = 262)

Matched pairs (n) CGM rate of change (mg/(dL·min)) %20/20 (%) MARD % (SE)
634 <−2 89.7 10.1 (0.4)
2489 [−2, −1) 93.0 9.6 (0.2)
8427 [−1, 0) 93.9 10.0 (0.1)
5741 [0, 1) 92.7 9.9 (0.1)
2049 [1, 2] 89.6 9.9 (0.2)
1624 >2 87.1 10.8 (0.2)