Table 1.

Analyte dilution accuracy and precision in human plasma (n = 4).

Analyte Theoretical Concentration (ng/mL) Experimental Concentration (ng/mL) %RE* %RSD*
HGA 5.00 × 102 5.06 × 102 1.2 4.6
1.00 × 103 1.01 × 103 0.80 4.6
5.00 × 103 4.60 × 103 7.9 1.9
1.00 × 104 9.97 × 103 0.35 4.6

MCPG 5.00 × 102 5.26 × 102 5.2 4.2
1.00 × 103 1.04 × 103 4.4 3.7
5.00 × 103 4.77 × 103 4.7 5.2
1.00 × 104 1.03 × 104 2.7 4.7
*

Percent relative error, ( %RE=[(Ce-Ct)Ct]×100%); Percent relative standard deviation, ( %RSD=[SDCavg]×100%)