Assay performance data of abiraterone in human plasma at validated quality control concentration levels

Nominal
conc.
(ng/mL)
Within-run
precision (%CV)
Within-run
accuracy
(% dev.)
Mean between run
precision (%CV)
Between-
run
Accuracy
(% dev.)
1 4.1 96.7 2.9 100.5
5 13.4 103.2 <13.4 101.2
300 2.2 95.2 2.0 97.6
400 2.9 96.5 1.4 98.2
500 4.5 95.4 2.8 98.9