Table 3:

Stability data of abiraterone

Nominal conc.
(ng/mL)
Determined mean
(ng/mL) (n=3)
Mean accuracy
% (n=3)
Stock solution 1 mg/mL (− 40°C, 7 months) 1.03 103
Quality control stability (− 40°C, 6 months)
5.00 4.7
93.2
300 308.3 103.0
400 410.0 102.7
Plasma 3 freeze-thaw (F/T) cycles
5.00 5.0 100.4
300
299.1 99.7
400 395.9 98.9
Plasma, (2-8°C), 8 days
5.00 5.4 107.0
300 298.4 99.5
400 388.0 97.0
Plasma, ambient temp, 8 days
5.00 5.2 103.5
300 298.5 99.5
400 396.2 99.1
Extract, autosampler (2-8°C), 8 days
5.00 5.1
102.5
300 299.0 99.7
400 405.8 101.5
Patient samples short term stability (2-8ºC) (% of baseline measurement)
Plasma 24hr
Whole blood 24hr
95.5
99.0
Patient samples short term stability (ambient temperature) (% of baseline measurement)
Whole blood 2hr
Whole blood 4 hr
Plasma 2hr
Plasma 3hr
Plasma 4hr
83.0
76.0
86.0
81.5
75.0