Table 1.
Manufacturer | Name | Assay principle | Calibration | Isoform-specificity | Measuring range | Recommended sample material | Comment |
---|---|---|---|---|---|---|---|
Siemens | Immulite 2000 | Two-site Chemilumminescent immunometric assay | 98/574 | Not provided | 0.05 to 40 ng/mL Analytical sensitivity: 0.01 ng/mL | Serum | ng/mL × 3.0 ➔mIU/L |
Diasorin | Liaison hGH | Chemiluminescent sandwich immunoassay | 98/574 | Not provided | 0.009–80 ng/mL | Serum | |
Beckmann-Coulter | Access Ultrasensitive hGH | Automated immunometric assay, Chemiluminescence | 98/574 | See comment | 0.002–35 ng/mL (μ/L) | Serum or plasma (heparin) | Cross reaction analysed with GH 8 ng/mL for 20 K-GH: − 2542% |
IDS | iSYS hGH | Chemilumminescent assay | 98/574 | 22kD GH: 100% | 0.05–100 ng/mL | Serum or plasma (heparin or EDTA) | Do not cross react with substances in these concentrations: 20 K-GH (10 ng/mL); placental GH (200 ng/mL); HPL (10,000 ng/mL); prolactin (40.000 ng/mL); pegvisomant (50,000 ng/mL); biotin 300 nmol/L); GHBP: 140 ng/mL. ng/mL × 3.0=μIU/mL |
DIASource | hGH IRMA | Immunometric assay enzyme amplified sensitivity | 98/574 | Not provided | 1–120 ng/mL Sensitivity: ng/mL |
Serum, plasma | Conversion factor: 1μIU = 0.33 ng |
hGH EASIA | Enzyme-Immunoassay | 98/574 | Not provided | 0.45–98 ng/mL Sensitivity:0.17 ng/mL |
Serum, plasma | Conversion factor: 1μIU = 0.33 ng | |
CisBio | hGH-RIACT | Immunoradiometric | 98/574 | Not provided | 0.03–75 ng/mL | Serum | 1 ng = 3μIU; do not cross react with prolactin and hPL. Cross reaction with 20kD GH is less than 5% for the concentration up to 3750 ng/mL (22kD GH proportional concentration above 24,000 ng/mL) |