Published in final edited form as: J Immunol Methods. 2017 May 18;448:44–50. doi: 10.1016/j.jim.2017.05.006

Fig. 1.

Fig. 1.

Validation Phase I and II data scatter plots of pairwise comparisons between FACSCalibur™ and the Pima™ point-of-care device for CD4+ T-lymphocyte counts of HIV-infected persons using venous and capillary blood samples. The solid and dashed lines are the 45 degree reference lines (perfect fit) and fitted lines from the linear regression model, respectively.