Published in final edited form as: J Immunol Methods. 2017 May 18;448:44–50. doi: 10.1016/j.jim.2017.05.006

Table 2.

Comparative performance of capillary and venous CD4+ T-lymphocyte counts on Pima™ POC and FACSCalibur™ by laboratory technicians and lay-counselors in Western Kenya.

Statistic FACS Calibur Venous CD4 Count FACS Calibur Capillary CD4 Count PIMA Lab Venous CD4 Count PIMA Lab Capillary CD4 Count PIMA lay-counselors Venous CD4 Count PIMA lay-counselors Capillary CD4 Count
FACS Calibur Venous CD4 Count Accuracy (Ca) 1.0 0.993 0.950 0.953 0.915 0.941
Precision (r) 1.0 0.968 0.960 0.963 0.954 0.958
CCC (rc) 1.0 0.962
0.947, 0.972
0.912
0.885, 0.932
0.918
0.893, 0.937
0.873
0.838, 0.900
0.902
0.873, 0.925
Bias 0.0 −26.9
−37.5, −16.4
−65.7
−78.4, −53.0
−63.0
−75.3, −50.6
−86.4
−100.4, −72.4
−71.2
−82.3, −58.2
FACS Calibur Capillary CD4 Count Accuracy (Ca) 1.0 0.979 0.981 0.953 0.973
Precision (r) 1.0 0.944 0.961 0.947 0.956
CCC (rc) 1.0 0.924
0.898, 0.943
0.943
0.924, 0.957
0.902
0.872, 0.926
0.930
0.908, 0.948
Bias 0.0 −38.8
−52.2, −25.3
−36.0
−47.5, −24.6
−59.5
−73.0, −45.9
−44.3
−56.5, −32.1
PIMA Lab Venous CD4 Count Accuracy (Ca) 1.0 0.9999 0.994 0.9996
Precision (r) 1.0 0.963 0.956 0.956
CCC (rc) 1.0 0.963
0.949, 0.973
0.950
0.932, 0.964
0.955
0.939, 0.968
Bias 0.0 2.7
−6.9, 12.4
−20.7
−31.2, −10.3
−5.5
−16.0, 5.0
PIMA Lab Capillary CD4 Count Accuracy (Ca) 1.0 0.993 0.999
Precision (r) 1.0 0.963 0.970
CCC (rc) 1.0 0.956
0.939, 0.968
0.970
0.958, 0.978
Bias 0.0 −23.5
−33.1, −13.9
−8.3
−16.9, 0.4
PIMA lay-counselors Venous CD4 Count Accuracy (Ca) 1.0 0.997
Precision (r) 1.0 0.957
CCC (rc) 1.0 0.954
0.937, 0.967
Bias 0.0 15.2
5.0, 25.4
PIMA lay-counselors Capillary CD4 Count Accuracy (Ca) 1.0
Precision (r) 1.0
CCC (rc) 1.0
Bias 0.0

Note to Table 2: CCC: concordance correlation coefficient. Bias (in units of cells/μl) was estimated using the Bland-Altman method. All comparisons to FACSCalibur™venous for CD4 counts are comparing to the reference. Comparisons for non-FACSCalibur™ venous have neither as a reference.