Figure 2.

Figure 2

Application accuracy results (n = 18 users). Correlation and Bland-Altman plots comparing the brachial BP measurements from the (ad) iPhone application and (eh) a finger cuff device to the brachial BP measurements from a standard arm cuff device. The closed circles are data from new users holding the finger devices at heart level, and the open circles are data from experienced users holding the finger devices below the heart to increase BP. r, correlation coefficient; μ, mean of errors (bias error); σ, standard deviation of errors (precision error).