Table 2.

Wide field IEC symmetry as measured with EPID and IC Profiler for all four available photon beams

Beam Plane IC Profiler symmetry (%) EPID symmetry (%) % difference
6 MV In‐plane 100.4 100.46 −0.06
Cross‐plane 100.3 100.34 −0.04
10 MV In‐plane 100.6 100.38 0.22
Cross‐plane 100.5 100.38 0.12
6 MV FFF In‐plane 100.4 100.23 0.17
Cross‐plane 100.4 100.00 0.40
10 MV FFF In‐plane 100.7 100.40 0.30
Cross‐plane 100.4 100.00 0.30
Mean difference 0.19 ± 0.18% (1 SD)