Table 1.

Summary results of monthly image quality tests (spatial resolutions, CNR, and uniformity) with the manufacturer specification for each test

[Range] or average ± SD
CTDI 2.5 mGy (low‐dose preset) CTDI 6.3 mGy (high‐quality preset) Manufacturer specification
Spatial resolution [6,7] lp/cm [7,8] lp/cm Minimum 6 lp/cm
CNR 1.07 ± 0.08 1.69 ± 0.10 >0.5 (CTDI 2.5)
>0.8 (CTDI 6.3)
Uniformity 12.84% ± 0.70 13.01% ± 0.69 <14%