Table 2. Diagnostic performance of the CareStart malaria (HRP2 and HRP2/pLDH) and SD bioline malaria (HRP2/pLDH) RDT kit to microscopy.

Gold standard Positive (N = 281) Gold standard Negative (N = 172) Sensitivity
% (95% CI)
Specificity
% (95% CI)
PPV
% (95% CI)
NPV
% (95% CI)
AUC
(95% CI)
RDT Positive (n) RDT
Negative (n)
RDT Positive (n) RDT
Negative (n)
Type of Malaria RDT
CareStart (HRP2) 276 5 58 114 98.2 (95.9, 99.4) 66.3 (58.7, 73.3) 82.6 (78.1, 86.5) 95.8 (90.5, 98.6) 0.82 (0.79, 0.86)
CareStart (HRP2/pLDH) 276 5 58 114 98.2 (95.9, 99.4) 66.3 (58.7, 73.3) 82.6 (78.1, 86.5) 95.8 (90.5, 98.6) 0.82 (0.79, 0.86)
SD Bioline (HRP2/pLDH) 276 5 53 119 98.2 (95.9, 99.4) 69.2 (61.7, 76.0) 83.9 (79.5, 87.7) 96.0 (90.8, 98.7) 0.84 (0.80, 0.87)