Skip to main content
. 2018 Mar 30;26(6):839–844. doi: 10.1016/j.jsps.2018.03.015

Table 4.

Robustness of the method by changing flow rate of the mobile phase and detection wavelength (n = 6).

Vitamin C
Gallic acid
Rutin
Ellagic acid
Retention time (RSD)% Peak area RSD (%) Retention time (RSD)% Peak area RSD (%) Retention time (RSD)% Peak area RSD (%) Retention time (RSD)% Peak area RSD (%)
Flow rate (mL/min) 0.8 0.88 0.44 0.28 0.38 0.78 0.36 0.58 0.88
1 1.02 0.53 0.87 0.58 1.25 2.11 0.40 0.53
1.2 0.45 1.58 0.66 1.55 1.23 0.52 0.87 0.74



Detection wavelength (nm) 250 0.84 1.46 0.98 0.84 0.84 0.78 0.96 0.64
254 0.71 1.28 1.11 0.57 0.56 0.56 0.28 0.66
258 1.58 0.35 0.58 0.71 0.38 0.44 1.62 1.84