Conversion coefficient of collimators: European vs. Japanese studies

Europe Japan* P values between Europe and Japan
N Mean SD N Mean SD
LEHR 103 0.553 0.018 73 0.552 0.048 0.85
LEGP+LEAP group** 10 0.591 0.024 25 0.648 0.036 <0.0001
 LEGP 4 0.570 0.011 17 0.654 0.037 0.0003
 LEAP 6 0.605 0.020 2 0.624 0.014 0.15
LMEGP 16 0.824 0.035 46 0.829 0.055 0.74
MEGP+MELP group 81 0.891 0.025 53 0.895 0.061 0.60
 MEGP*** 28 0.882 0.017 40 0.878 0.054 0.71
 MELP 53 0.895 0.027 13 0.950 0.051 <0.0001

*Data from J Nucl Cardiol 2014; 21: 970–978

**LE general-all-purpose collimator is included in Japanese study

*** MEGP, ME general-all purpose, and ME collimators are included in Japanese study