Table 2.

Laser parameters set during SMILE in various treatment modes

Laser parameter Expert modea Standard modeb Fast modec
Pulse energy 140–200 nJ 130 nJ 170 nJ
Energy offset (1 offset = 5 nJ) 28–40 26 34
Track distance Lenticule and cap cuts: 4.5 ± 0.5 µm Lenticule and cap cuts: 3 µm Lenticule and cap cuts: 4.5 µm
Lenticule and cap side cuts: 2 ± 0.5 µm Lenticule and cap side cuts: 2 µm Lenticule and cap side cuts: 2 µm
Spot distance Lenticule and cap cuts: 4.5 ± 0.5 µm Lenticule and cap cuts: 3 µm Lenticule and cap cuts: 4.5 µm
Lenticule and cap side cuts: 2 ± 0.5 µm Lenticule and cap side cuts: 2 µm Lenticule and cap side cuts: 2 µm

Notes:

a

The expert mode allows the surgeon to vary the “fast” mode settings within a certain range.

b

The standard settings are preset by the manufacturer.

c

The fast settings are set by the application specialist and may vary from surgeon to surgeon.

Abbreviation: SMILE, small incision lenticule extraction.