Table 5.

Cost-Utility Results.

Costs (2016 pound sterling) QALYs Combined hypoglycemic eventsa Incremental cost-effectiveness ratio (pound sterling per QALY)
Scenario 1 (lifetime time horizon). % error associated with HbA1c but NOT directly associated with hypoglycemic events
Type 1 diabetes
SMBG device with 15% error, £0.20 per strip £81 809 15.292
SMBG device with 8.4% error, £0.30 per strip £82 182 15.414 £3064/QALY
Type 2 Diabetes
SMBG device with 15% error, £0.20 per strip £39 523 9.743
SMBG device with 8.4% error, £0.30 per strip £41 176 9.750 £264 668/QALY
Scenario 2 (3-year time horizon). % error associated with hypoglycemic events but NOT directly associated with changes in HbA1c
Type 1 diabetes
SMBG device with 15% error, £0.20 per strip £7296 2.142 4.85
SMBG device with 8.4% error, £0.30 per strip £7191 2.155 3.74 Dominantb
Type 2 diabetes
SMBG device with 15% error, £0.20 per strip £5223 2.166 4.25
SMBG device with 8.4% error, £0.30 per strip £5174 2.178 3.28 Dominantb
a

All severe hypoglycemic events requiring inpatient visits and hypoglycemic events requiring outpatient visits.

b

Less costly and more effective.