. 2018 Nov 23;3:110. Originally published 2018 Sep 4. [Version 2] doi: 10.12688/wellcomeopenres.14767.2

Table 3. Comparison of Bilistick System (BS) reader#1 and #2 from available results (n=82).

Variables * All (n=82) Reader #1
(n=36)
Reader #2
(n=46)
p-value
Newborn characteristics
Gestational age, weeks 39 [38-40] 39 [38-40] 39 [38-40] 0.359
Age, hours 127 [70-169] 127 [82-221] 125 [68-168] 0.824
Haematocrit, % 53 [49-55] 50 [49-54] 54 [52-56] 0.046
Environment characteristics
Room temperature, °C 29 [28-30] 28 [27-31] 29 [29-30] 0.782
Room humidity, % 69 [58-78] 58 [52-65] 76 [69-81] <0.001
Humidity level, n (%) <0.001
Normal 52 (63.4) 33 (91.7) 19 (41.3)
High 30 (36.6) 3 (8.3) 27 (58.7)
Bilirubin measurement characteristics
BS bilirubin, µmol/l 168 [138-218] 187 [136-226] 164 [138-186] 0.266
Bilirubinometer bilirubin, µmol/l 192 [150-229] 214 [151-251] 181 [150-205] 0.045
PCC of bilirubin values between BS and
routine laboratory bilirubinometer, r (95% CI)
0.87 (0.77-0.93) 0.97 (0.93 - 0.98) 0.71 (0.53 - 0.83)

*Data presented as median [Interquartile range] unless otherwise stated. PCC, Perason’s correlation coefficient.