Skip to main content
. Author manuscript; available in PMC: 2018 Sep 14.
Published in final edited form as: J Orthop Trauma. 2017 Jul;31(7):363–368. doi: 10.1097/BOT.0000000000000836

Fig 4.

Fig 4.

The axial stiffness of each group. The y-axis is the axial stiffness in N/mm. The cannulated screw groups were not significantly different from each other (p=0.9222). The CHS group was significantly more stiff than either cannulated screw group (CHS vs CS p= 0.0005; CHS vs IT p=0.0005).