Table 1.

Detailed sequence parameters of the acquired cine CMR sequences.

1.5 T
3 T
7 T
FLASH TrueFISP FLASH TrueFISP FLASH
TR [ms] 49.14 39.39 50.76 29.1 40.9
TE [ms] 3.45 1.36 2.89 1.27 4.76
Matrix [pixel] 208*136 208*170 208*141 208*144 240*196
Field of View [mm2] 340*278 340*278 340*278 340*285 360*294
Flip angle [°] 15 54 12 38 70
Segments 7 13 9 10 5
Calculated phases 25 25 25 25 25
Spatial resolution [mm3] 1.6 × 1.97 × 4 1.6 × 1.6 × 4 1.6 × 1.97 × 4 1.6 × 1.98 × 4 1.5 × 1.5 × 3
Voxel Volume [mm3] 12.6 10.2 12.6 12.7 6.8
Temporal resolution [ms]
(given a mean heart rate of 70 bpm)
2.8 2.8 2.8 2.8 2.8
Bandwidth [Hz/px] 253 925 445 1502 992
pMRI Grappa R = 1 R = 2 R = 1 R = 3 R = 2
Number of active RF coil channels 20-32 Rxa 20-32 Rxa 20-32 Rxa 20-32 Rxa 8 Tx/Rx

FLASH, Fast Low-Angle Shot sequence; TrueFISP, True Fast Imaging balanced Steady-state free Precession; TR, repetition time; TE, echo time; pMRI, parallel MRI; RF, radiofrequency; Rx, receive; Tx, transmit;

a

Using auto coil select mode. In TrueFISP relatively low flip angle had to been chosen due to SAR-restrictions.