Skip to main content
. 2018 Sep 11;5:159–164. doi: 10.1016/j.ejro.2018.08.008

Table 2.

Qualitative artefact severity (given as median (range)).

Sequence/ field strength reader #1 reader #2
7 T FLASH 0 (0–1) 1 (0–1)
3 T FLASH 1 (0–1) 1 (0–1)
1.5 T FLASH 1 (0–1) 1 (1–2)
3 T TrueFISP 2 (1–4) 2 (1–3)
1.5 T TrueFISP 2 (0–4) 2 (0–3)

0, no artefacts; 1, slight artefacts; 2, considerable artefacts; 3, severe artefacts; 4, image analysis not feasible.